Contact person: Mr.Hu
Tel: 0411-87300505
Fax:0411-87305556
Mobile: 13940906200
E-mail: fenghu@chjtech.com
Factory Address:
NO.5,HuaiheRiverthreegroups,
Economicdevelopmentzone,Dalian
|
|
|
|
|
| Rotary Evaporator,RE300/600/800 |
| |
| |
   Index    |
| Features |
function (RE800) |
| 53 Solvents in the memory of RE800 |
The Saturated Vapor Pressure Curve In The Memory |
| Configuration of power plugs |
Specifications |
| Specification of optional Vacuum Regulator VR300 |
Specification of optional Vapor Temperature Indicator TA300 |
| Models for your choice |
Nomenclature by model |
| Optional accessories |
Peripheral equipment |
Features |
| 1.Motorized lift standard for all models |
2.Digital settings and display in Japanese or English |
 Yamato Rotary Evaporators, models RE300/600/800, are easily raised or lowered by simple operation of switch. Model RE800, can be pre-set, so that when operation of the rotary evaporator ceases, unit will be raised. |
 The evaporator's motor speed(rpm), vacuum(degrees) and vapor temperature can be digitally set and displayed. Experimental conditions and steps during a reaction can be saved and repeated in future. Display language can be chose either Japanese or English.(Only the rpm of the motor can be digitally displayed in model RE300.)
|
|
3.Space-saving design
|
4.Movable rotary joint
|
 The rotary evaporator is compact, which can be neatly installed in a fume hood. Despite its depth of 400 mm, the vertical condenser (Glassware B/C) does not take up space. |
 The rotary joint's locking position feature is adjustable within 80 mm; especially useful when using varying capacity evaporating flasks or when there is a need to shift the vacuum seal's contact position. (Patent No. 3220033) |
| 5.Quick release of evaporating flask |
6.Newly designed glass components (Guard against liquid stagnation) |
 The evaporating flask can be released easily and quickly by just turning the flask release nut. |
 Vertical condenser(Glassware B&C) function to prevent liquid from stagnating. (patent pending) New condenser tubes (type A&B) designed to increase rapidity of cooling capacity.(Surface area is 20% larger than previous model.) |
7.Water bath (oil bath) |
8.Voltage-sensing power pack |
 Original removable bath enables easy cleaning and replacing of the water(oil). The digital temperature indicator is designed for easy setting and reading. The most suitable bath is selectable from model BM Water Bath or Model BO Oil Bath in accordance with working temperatures of the experiment. |
 The rotary evaporator (including the water bath, oil bath and some optional accessories) works with a power supply of 100-240V AC. For use overseas, please choose a plug according to shape of socket your abode among configuration of power plugs(and cord) as an option herein. |
| 9.Hose joints |
10.Others |
 Resin-made hose joints are used for easy connection of cooling water and vacuum hoses to their corresponding glass ports. |
*Vacuum seal The fluoro-rubber vacuum seal with its multistage configuration is designed to maintain the stability of reduced pressures. Pure Teflon vacuum seal is also available as an option for evaporation of ether-based and ketone-based liquids.。
*Carefully designed safety measures The evaporator can detect any malfunctions in the driving mechanisms in the rotary section and the lifting device, and will automatically stop to ensure safety. It can also automatically detect abnormal temperatures in the water bath and oil bath and will stop the operation of the evaporator(only for models RE600/800).
*Stable rotation at low and high speeds Together with the digital setting and display, the rotary evaporator maintains a stable rotation by using the feedback control feature. This ensures precise reproducibility of experiments.
|
RE300
|
Motorized lift
Feedback control motor
Voltage-sensing power supply
Digital setting and rpm display
Stable rotation |
 RE300A-W |
 RE300B-W |
 RE300C-W |
function (RE800) |
| -Vacuum regulation - |
| Boiling point indicates a temperature at which ambient pressure becomes equal to the saturated vapor pressure of a liquid. The saturated vapor pressure varies with the temperature. The pressure goes down when the temperature decreases. "Vacuum regulation" is the reduction in the internal pressure of a vessel which can decrease the boiling point of a liquid. |
| -Vapor temperature - |
| Once a liquid reaches its boiling point, it starts vaporizing, generating bubbles and foam. (This is different from what is generally known as "bumping".) The temperature of the vapor in this state is called "vapor temperature". |
| -Automatic distillation- |
| After the desired vapor temperature is preset, pressure is reduced until the temperature reaches the preset temperature. The reduced pressure is maintained at this temperature. When the vapor temperature deviates from its setting, the automatic distillation feature will either proceed to the next step (setting) in the operation or stop the operation there. This feature detects even a fine fluctuation of pressure and automatically regulates it to prevent bumping and ensure gradual change in pressure. |
53 Solvents in the memory of RE800 |
For Low Boiling Point
|
1,1-Dichloroethane
|
|
1,2Dichloroethylene cis
|
|
1,2Dichloroethylene trs
|
|
1,2-Dichloroethane
|
| 1-Propanol |
|
2-Butanol
|
| 2-Propanol |
| n-Hexane |
|
n-Pentane
|
| Acetone |
|
Ethyl Acetate
|
| Ethanol |
| Chloroform |
| Ethyl acetate |
| Methyl acetate |
|
Diisopropyl Ether
|
|
Diethyl Ether
|
| Carbontetrachloride |
| Cyclohexane |
| Cyclopentane |
|
Dichloromethane
|
| Ethylbromide |
|
Heptane
|
| Benzene |
| Pentane |
|
Pentachloroethane
|
| Water |
|
Methanol
|
|
For High Boiling Point
|
1,1,2-Trichloroethane
|
|
1-Octanol
|
|
1-Nonanal
|
|
1-Butanol
|
|
1-Hexanol
|
|
1-Heptanol
|
|
1-Pentanol
|
|
2-Pentanol
|
|
N,N-Dimethylformamide
|
|
m-Xylene
|
|
o-Xylene
|
|
p-Xylene
|
|
Isoamyl Alcohol
|
|
Ethylene Glycol
|
|
Hexanoic acid
|
|
Formic acid
|
|
Glycerol
|
|
Chlorobenzene
|
| Acetic acid |
|
Propyl acetate
|
|
Dioxane
|
|
Styrene
|
|
Tetrachloroethylene
|
| Toluene |
|
Butanic acid
|
|
The Saturated Vapor Pressure Curve In The Memory |
 |
|
The following Antoine Formula is often used for the calculation of liquid’s steam pressure. log P = A - B / ( C - t )・・・・・Antoine(Antoine’s Formula)P : Steam Pressure(kPa) T : Temperature(deg.C), A, B, C are called “Antoine Constant” and they are various depending on the substance.
|
| |
Country |
Voltage |
Frequency |
Configuration of plug |
Asia Middle East |
Iran |
220/230V |
50Hs |
C-2,BF |
| India |
220/230/240V |
50Hs |
C-2,B-3,BF |
| Indonesia |
127/220/230V |
50Hs |
A-2,C-2,B-3,BF |
| Korea |
100/200/220V |
60Hs |
A-2,C-2 |
| Saudi Arabia |
127/220V |
50/60Hs |
A-2,C-2,B-3,BF |
| Singapore |
110/230V |
50Hs |
B-3,BF |
| Thailand |
220/240V |
50Hs |
C-2,BF |
| Taiwan |
110/220V |
60Hs |
A-2,C-2,O |
| China |
110/220V |
50Hs |
A-2,C-2,B-3,O,BF |
| Pakistan |
220/230V |
50Hs |
C-2,B-3 |
| Philippine |
110/220/230/240V |
60Hs |
A-2,C-2,B-3,O,BF |
| Vietnam |
110/220V |
50Hs |
A-2,C-2 |
| Hong Kong |
110/220V |
50Hs |
C-2,B-3,BF |
| Malaysia |
240V |
50Hs |
C-2,B-3,BF |
| North America |
USA |
120V |
60Hs |
A-2 |
| Canada |
120/240V |
60Hs |
A-2,BF |
| Hawaii |
110/120V |
60Hs |
A-2 |
| Mexico |
110/120/220V |
60Hs |
A-2 |
| South America |
Argentina |
220V |
50Hs |
C-2,O,BF |
| Colombia |
110/120/150V |
50Hs |
A-2 |
| dust |
220V |
50Hs |
C-2,B-3 |
| Brazil |
127/220V |
60Hs |
A-2,C-2 |
| Peru |
220V |
60Hs |
C-2 |
| Africa |
Egypt |
220V |
50Hs |
C-2,B-3,BF |
| Kenya |
240V |
50Hs |
C-2,B-3,BF |
| South Africa |
220/230/250V |
50Hs |
C-2,B-3,BF |
Specifications |
| Model |
RE300 |
| Composition |
Motor |
DC brushless motor (for rotation) and DC motor (for lifting) |
Performance and features |
Number of revolutions |
10 - 250 r/min (rpm) |
| Motorized lift stroke |
150mm |
| Speed (rpm) setting |
Control knob (digital indication) |
| Safety measures (main unit) |
Manual-setting lower limit, upper limit, motor overload |
| Lifting feature |
Motorized lifting system |
| Specifications |
External dimension *1 |
With Glassware A (mm) |
828(W)× 400(D)×580(H)(740(H)when raised) |
| With Glassware B/C (mm) |
643(W)× 400(D)×717(H)(877(H)when raised) |
| Weight |
Approx. 13kg(excluding the glassware and the water/oil bath) |
| Power supply |
Main unit |
100 - 240V AC, 1.5A(excluding the water/oil bath) |
| |
Model |
RE600 |
RE800 |
| Composition |
Motor |
DC brushless motor (for rotation) and DC motor (for lifting) |
| Controller |
Vacuum regulator VR600 |
Vacuum regulator VR800 |
| Performance |
Number of revolutions |
10 - 250 r/min (rpm) |
| Motorized lift stroke |
150mm |
| Setting range of vacuum |
0-981hPa |
| Measurable range of vacuum |
0-1033hPa |
| Resolution of vacuum |
1hPa |
| Setting range of hysteresis |
1-50hPa |
| Resolution of vapor temperature indicator |
Selectable(either 1°C or 0.1°C) |
| Readout of cooling water temperature |
Depending on indicator(option) |
| Resolution of cooling water temperature indicator |
1°C |
Performance and features |
Operation modes |
Manual without vacuum control, preset, preset with timer, pressure gradient, and gradient with timer |
| - |
Auto I (drying), Auto II (for one solvent), and Automatic III (for more than two solvents) modes |
| Setting range of timer |
1-999 minutes in increments/decrements of 1 minute for preset operations,1-99 minutes for progressive operations |
| Memory |
10 programs/operation mode |
| Data of solvents |
- |
53 saturated vapor pressure curves |
| Speed (rpm) setting |
Control knob (digital indication) |
| Safety measures (drive unit) |
Manual-setting lower limit, upper limit, motor overload |
| Safety measures (vacuum regulator) |
Self-diagnosis, main unit/bath synchronized stop at malfunctions |
| Synchronized control feature |
Selection of automatic bath stop or automatic insulation |
| Lifting feature |
Motorized lifting system |
Motorized lifting system synchronized with control |
| Specifications |
External *1 dimension |
With Glassware A (mm) |
828(W)× 400(D)×721(H)(881(H)when raised) |
| With Glassware B/C (mm) |
643(W)× 400(D)×721(H)(881(H)when raised) |
| Weight |
Approx. 14kg(excluding the glassware and the water/oil bath) |
| Power supply |
Main unit |
100 - 240V AC, 1.5A(excluding the water/oil bath) |
※1 External dimensions (excluding protrusions) ※2 A microprocessor memorizes the saturated vapor pressure of each solvent and automatically presets an optimum vacuum pressure. |
Specification of optional Vacuum Regulator VR300 |
| Vacuum sensor |
0 to 981hPa |
| Setting range of vacuum |
0-981hPa |
| Measurable range of vacuum |
0-1033hPa |
| Resolution of vacuum |
1hPa |
| Setting range of hysteresis |
1-50hPa |
| Operation modes |
Customer-programmed, presetting, presetting with timer, progressive, and progressive with timer modes |
Setting range of timer |
Preset operation |
1 - 999minutes |
| Progressive operation |
1 - 99minutes |
| Safety measures |
Self-diagnosis, alarm |
Specification of optional Vapor Temperature Indicator TA300 |
| Vapor temperature sensor |
Pt100 ohem |
| Steam temperature indicator |
Digital indication |
| Resolution of vapor temperature indicator |
0.1°C |
Models for your choice |
| Main unit |
Product code |
Model |
Glassware |
Bath(100-120V AC) |
Vacuum regulator |
Vapor temperature indicator |
| Water bath |
Oil bath |
| A |
B |
C |
BM500 |
BO400 |
VR |
TA300 |
| RE300 |
255370 |
RE300A |
 |
|
|
|
|
|
|
| 255113 |
RE300A-W |
 |
|
|
 |
|
|
|
| 255116 |
RE300A-O |
 |
|
|
|
 |
|
|
| 255371 |
RE300A-V |
 |
|
|
|
|
 |
|
| 255372 |
RE300A-WV |
 |
|
|
 |
|
 |
|
| 255373 |
RE300A-OV |
 |
|
|
|
 |
 |
|
| 255374 |
RE300A-T |
 |
|
|
|
|
|
 |
| 255375 |
RE300A-WT |
 |
|
|
 |
|
|
 |
| 255376 |
RE300A-OT |
 |
|
|
|
 |
|
 |
| 255380 |
RE300B |
|
 |
|
|
|
|
|
| 255114 |
RE300B-W |
|
 |
|
 |
|
|
|
| 255117 |
RE300B-O |
|
 |
|
|
 |
|
|
| 255381 |
RE300B-V |
|
 |
|
|
|
 |
|
| 255382 |
RE300B-WV |
|
 |
|
 |
|
 |
|
| 255383 |
RE300B-OV |
|
 |
|
|
 |
 |
|
| 255384 |
RE300B-T |
|
 |
|
|
|
|
 |
| 255385 |
RE300B-WT |
|
 |
|
 |
|
|
 |
| 255386 |
RE300B-OT |
|
 |
|
|
 |
|
 |
| 255390 |
RE300C |
|
|
 |
|
|
|
|
| 255115 |
RE300C-W |
|
|
 |
 |
|
|
|
| 255118 |
RE300C-O |
|
|
 |
|
 |
|
|
| 255391 |
RE300C-V |
|
|
 |
|
|
 |
|
| 255392 |
RE300C-WV |
|
|
 |
 |
|
 |
|
| 255393 |
RE300C-OV |
|
|
 |
|
 |
 |
|
| 255394 |
RE300C-T |
|
|
 |
|
|
|
 |
| 255395 |
RE300C-WT |
|
|
 |
 |
|
|
 |
| 255396 |
RE300C-OT |
|
|
 |
|
 |
|
 |
| RE600 |
255123 |
RE600A |
 |
|
|
|
|
Standard equipment |
| 255130 |
RE600A-W |
 |
|
|
 |
|
Standard equipment |
| 255133 |
RE600A-O |
 |
|
|
|
 |
Standard equipment |
| 255124 |
RE600B |
|
 |
|
|
|
Standard equipment |
| 255131 |
RE600B-W |
|
 |
|
 |
|
Standard equipment |
| 255134 |
RE600B-O |
|
 |
|
|
 |
Standard equipment |
| 255126 |
RE600C |
|
|
 |
|
|
Standard equipment |
| 255132 |
RE600C-W |
|
|
 |
 |
|
Standard equipment |
| 255135 |
RE600C-O |
|
|
 |
|
 |
Standard equipment |
| RE800 |
255148 |
RE800A |
 |
|
|
|
|
Standard equipment |
| 255140 |
RE800A-W |
 |
|
|
 |
|
Standard equipment |
| 255143 |
RE800A-O |
 |
|
|
|
 |
Standard equipment |
| 255149 |
RE800B |
|
 |
|
|
|
Standard equipment |
| 255141 |
RE800B-W |
|
 |
|
 |
|
Standard equipment |
| 255144 |
RE800B-O |
|
 |
|
|
 |
Standard equipment |
| 255150 |
RE800C |
|
|
 |
|
|
Standard equipment |
| 255142 |
RE800C-W |
|
|
 |
 |
|
Standard equipment |
| 255145 |
RE800C-O |
|
|
 |
|
 |
Standard equipment |
Nomenclature by model |
 |
Optional accessories |
|
Teflon Solenoid valve (VR300/600/800)
|
Evaporation Flask 50ml
|
Transparent Nestle
|
|
VR300/600/800 SUS Pressure Sensor
|
Evaporation Flask 100ml
|
Transparent Nestle
|
|
VR800 Cooling Water Temp. Display Sensor
|
Evaporation Flask 200ml
|
Transparent Nestle
|
|
RE300/600/800Trap (With Horse Nozzle)
|
Evaporation Flask 300ml
|
Transparent Nestle
|
|
RE300/600/801Horse Connecting Parts
|
Evaporation Flask 500ml
|
Transparent Nestle
|
|
PG200 20L Teflon Diaphragm Dry Pump (AC100V)
|
Evaporation Flask 2000L
|
Transparent Nestle
|
|
VR600/800 Vacuum Pump Control Unit
|
Receiving Flask 300ml
|
Nestle
|
|
Cooling Horse φ9 2m×1pc.
|
Receiving Flask 500ml
|
Nestle
|
|
Vacuum Horse φ6 5m×1pc.
|
Receiving Flask 2L
|
Nestle
|
|
Triple Neck Cock
|
Joint T29/42 TransparentーT29/38 Transparent
|
|
Stand With Caster (155A With Receptacle)
|
Joint T29/42 Transparent―T24/40 Transparent Joint T29/42 Transparent―T19/38 Transparent Joint T29/42 Transparent―T15/25 Transparent Joint T24/40 Transparent―T24/40 Transparent Trap Ball T29/42 Transparent―T29/42 Transparent Trap Ball T24/40 Transparent―T29/42 Transparent
|
Peripheral equipment |
 |
RE800B+Vacuum pump DTC-21+Solvent recovery unit RT200+Cooling water circulation unit CF300+Cart |
 |
RE800B+Solvent recovery unit RT101 |
 |
RE800B+Vacuum pump DTC-21+Cooling water circulation unit CF300 |
| Product code |
Description |
Model |
Working temperature |
Cooling capacity |
Power supply |
| 255276 |
Solvent recovery unit |
RT101 |
-20°C - room temperature |
385kcal/h at 10°C |
100V AC |
| 255281 |
Solvent recovery unit |
RT200 |
- |
- |
100V AC |
| 221511 |
Cooling water circulation unit |
CF300 |
-20°C - room temperature |
385kcal/h at 10°C |
100V AC |
| 221513 |
Cooling water circulation unit |
CF700 |
-20°C - room temperature |
1,030kcal/h at 10°C |
100V AC |
| 221516 |
Cooling water circulation/pump unit |
CF700P |
-20°C - room temperature |
1,030kcal/h at 10°C |
100V AC |
| Product code |
Description |
Model |
Ultimate vacuum* |
Displacement |
Power supply |
| 242425 |
Handy aspirator |
WP25 |
2.3kPa at 20°C |
11-14ml/min×2 |
- |
| 242424 |
Handy aspirator |
WP15 |
2.3kPa at 20°C |
11-14ml/min×2 |
100V AC |
| 20UL41 |
Diaphragm vacuum pump |
DTC-41 |
1000Pa |
40/46ml/min(50/50Hz) |
100, 115, 200, 220,230V AC |
| 20UL21 |
Diaphragm vacuum pump |
DTC-21 |
1000Pa |
20/24ml/min(50/50Hz) |
100, 115, 200, 220,230V AC |
| Depending on the water vapor pressure at a given working temperature |
| Product code |
Description |
Model |
Attainable lowest temperature |
Trap capacity |
Dehumidification |
Power supply |
| 221511 |
Cooling trap |
CA300 |
-45°C |
Approx. 4L |
0.9kg |
100V AC |
| 221487 |
Glass condenser for CA300 |
- |
- |
- |
- |
- |
|
| |
|
|
|
|